Portable analyzer and processor of isokinetic parameters, able to measure and calculate with relation to physical quantities such as Temperature, Air speed, measurement of static and dynamic pressure and measurement of atmospheric pressure, directly calculating normalised air volume. Equipped with a double sensor for operating pressure detection to allow for further accuracy in ranges even with the lowest available minimum levels.
Equipped with a special area and drop number calculation, including for stacks with sections differing from standard circular stacks. The back-lit display shows all quantities in real time and their corresponding calculations for setting sampling devices and related accessories. All measurements and respective data can be memorised and downloaded on a PC by means of external software via RS232. The Isoflow 201 processor is interfaced with a sampler board (AirCube, AirFlow) and allows for use of the sampler in real time Isokinetic mode, modifying to change the sampler flow rate for emissions situation changes and permitting adjustments to sampling development. An easy, guided menu allows the operator to visualise all set and memorised data and to modify them even during the sampling phase.
Other features are:
• Measurement range, see following technical information
• Operating range -10/+40°C
• Pressure detection: double sensor with automatic scale change (see following Pro version technical information)
• Dimensions: 20x10x6 cm
• Weight: 950 grams
• Data output: RS232 bi-directional
• Rechargeable 12Vcc NiMH power
The Isoflow 201 measurer is supplied complete with: power feed/battery charger, case for transport and user manual in Italian, pipe fittings and cable for temperature sensor.
Electrical |
Power |
NiMH internal batteries
External power by means of DC (16<V<24) or AC (12<V<18) 1A source |
Memory |
8 kbytes 1M written |
Interface |
RS 232C 9600/38400 baud, 1stop, no parity |
Environmental Conditions |
Temperature |
Storage range: -20.. +70°C
Maximum Operating range: 0.. +50°C
Operation within specific precisions: 5.. +35°C |
Humidity |
Relative from 10% to 90% non condensing |
Mechanical |
Weight |
620 gr. |
Temperature sensor |
Measurement range |
Depends on type of probe and instrument settings |
Probe voltage reading range
(-->FS = 156,25 mV) |
+/- 78,125 mV |
Non linearity |
< 0,001% FS |
Resolution |
1.2μV [0,03°C for K probe under 150°C ] |
Precision |
0.07% on measurement (gain error) + 0.001% FS [approx 0.16°C for K probe with T < 150°C ] |
Conversion error
voltage --> temperature |
< 0.05°C |
Cold point |
Measurement range |
- 10.. + 85°C |
Resolution |
0.05°C |
Precision |
+/- 0.5°C |
Absolute sensor pressure |
Maximum range2 |
0.. 2065 mbar absolute (30 psi) on negative instrument input |
Reading range |
More than measurement range and dependent of sensor |
Measurement range3(FS) |
500.. 1100 mbar |
Resolution |
0.05 mbar (minimum: variable with sensor features) |
Precision |
< +/- 1.5 mbar (maximum with manufacturing calibration) |
Differential sensor pressure (Standard Version) |
Maximum range4 |
+/- 2500 mm H2O (+/-3,5 mmH2O diff.) |
Reading range |
+/- 2500 mm H2O |
Measurement range(FS) |
+/- 700 mm H2O (+/-1mmH2O diff. @ ambient T. of calibration given offset calibration) |
Resolution |
0.1mm H2O |
Precision5 |
+/- (0.15 (adc) + 3.5 (sensor) mm H2O minimum without calibration
with manufacturing calibration:
-0.9% on scale (FS) 0.. 100 mmH2O with reading less than 100 mm H2O
-0.8% on scale (FS) 0.. 300 mmH2O c with reading less than 300 mm H2O
-0.8% on scale (FS) 0.. 700 mmH2O with reading less than 700 mm H2O |
Non linearity + hysteresis |
+/- 0.5% FS |
Gravity effect6 |
+/- 1 mmH2O |
Differential sensor pressure (Pro Version) 0-100 mmH2O |
Measurement range(FS) |
0-100mm/H2O (+/- 1 mmH2O diff. on range 5..50°C given offset instrument calibration) |
Resolution |
> 0.01 mmH2O |
Precision |
1% FS on range 5..50°C (0,5% FS @ ambient T. of calibrationgiven offset calibration) |
Non linearity + hysteresis |
+/- 0.3% FS max (standard 0.15% FS) |
Gravity effect |
< +/- 0.1 |
Calibration |
Ambient T. |
20.. 25°C |
Instrument position |
horizontal +/- 20° |
| Product |
code |
Misuratore di Portata ISOFLOW 201 Standard
|
600/IF201ST |
| Misuratore di Portata ISOFLOW 201 Pro |
600/IF201PRO |
| Cavo di ricambio 1500mm per sonda temperatura |
600/RTC0020 |
| Software trasferimento/elaborazione dati per IF200 |
600/0SW200 |
| Valigia per trasporto IsoFlow ed accessori |
600/PPC1500BNF-IF |
| Confezione 10 Rotoli carta termica per stampante |
600/RCT57X50 |
| Cavo di ricambio 10 mt. per sonda temperatura |
600/RTC0010 |
| Cavo di ricambio 5 mt. per sonda temperatura |
600/RTC0025 |
| Stampante portatile per Isoflow e Serie Plus |
600/SPRINTB22 |
| Tubo al silicone 4x8 per IsoFlow200 1mt |
600/TS00048 |