Asbestos fibers sampling
|
|
|
|
Total particle sampling |
|
|
|
Sampling for PM10 and PM2.5 particles conforming to EN-12341 and EN-14907 |
|
|
|
Normalised Volume reading at display |
|
|
|
Reading of temperature and atmospheric pressure (optional for pressure) |
|
|
|
Set up program in time and volume of the sampling |
|
|
|
Manual adjustment of the flow with proportional valve and electronic compensation for loss of load |
|
|
|
Samplings settings by sector of origin of wind (with optional VDV sensor installed and interface) |
|
|
|
Memorisation of average sampling data such as volume, flow and temperature |
|
|
|
Visualisation of the yield of the flow and of the standard deviation with setting of tolerance parameters |
|
|
|
Optional interface with sequential groups for the sampling of total PM10 and PM2.5 particles TSP |
|
|
|
Interface with meteorological sensor group (optional) |
|
|
|
RS232 Interface for exhaust data and sample parameters |
|
|
|
Remote control with GSM/GPRS device (optional) |
The constant flow sampler AirCube HS is available with new features in its range of flow, in its new winning design and its ulterior practicality and versatility in its utilisation. In the new line the AirCube HS sampler shows a slightly reduced size for an increase in its already great handling and versatility, which has identified its range for years. The new AirCube HS line distinguishes itself also for its elevated range of accessories with which it may be equipped as much for environmental applications as for applications of sampling in emission. The model HS is equipped with an aspirating pump with graphite palettes, automatic by-pass, type 5 latest generation data card for control and management. The flow control comes about by way of the manual proportional valve and the compensation of the loss of load is by electronic control. The principal application is in the sampling of PM10 and PM2.5 particles under the conditions foreseen by the European norms and acknowledged by EN-12341 and EN-14907 to PM10 and PM2,5 particle fraction sampling. Obviously maintaining the dimensions similar to the other models of AirCube samplers but with the difference that the rotating pump is lodged in a separate part of the instrument bay allowing greater ventilation for cooling; it is utilisable for all sampling applications where there is foreseen an elevated flow of aspiration and also with elevated losses of load. The 5 data card permits the detection and visualisation of instantaneous temperature at the counter, the atmospheric barometric pressure (optional) the partial volume of litres sampled. Moreover the display visualises the volume of normalised air in function of the atmospheric pressure and the average temperature and pressure values the AirCube HS sampler memorises the sampling data even in case of the absence of electrical power supply. Upon the return of the same, the sampling begins again at the point of interruption.
Technical Specifications
-
Range of use 5-50 litres / minute
- Flow at free mouth 6m3 /hour
- Pump type Rotational with palettes
- Flow meter range Electronic set point on LCD Display
- Compensation Over 600mm/Hg
- Reading total litres Dry Gas volumetric counter
- Temperature control On display with precision of 0.1 ° C
- Cooling Forced air change every 5 seconds and aluminium dissipater
- Data Outlet RS 232 port for PC or portable serial printer connection
- Atmospheric Pressure control With Sensor installed directly on the data card (optional)
- Remote control With GSM/GPRS modem (optional)
- Sequential interface Optional
- Wind velocity control With VDV sensor (optional)
- External structure Painted aluminium alloy (Rohs 2002/95/EC)
- Dimensions: (LxDxH) 27x27x42 cm
- Weight 16 Kg.
- Electrical power feed: 220V-AC/50 Hz